Start/Stop by Digital Input (DI)

 

Applies to

Sensoft Vision

Issue

Where do I connect the wires for automatically start and stop measurements in Sensoft Vision?

Figure 1: Schematics of the DI hardware.

Hardware connection

You can use Digital Inputs (DI) to start and stop measurements. The required DI signal is a dc signal that goes to +24 V for starting the measurement and returns to ground when you want to stop the measurement. The Input configuration is by default Emitter-Follower, see Figure 1. Also when using digital inputs to start/stop measurements, you can still start/stop the measurements manually by clicking the Start/Stop button in Sensoft Vision.

Figure 2: Back panel of the Sensystem Network with connectors DI.

Sensystem Network with PXL

Connect the wires to the following pins of connector DI on the SenSystem Network, Figure 2:

  • Pin 5: Start/Stop signal
  • Pin 26: Ground

The connector that goes into connector DI of the SenSystem Network is a 37 pin male D-SUB connector. We offer a set with cable and screwable break-out.

Figure 3: PXT sensor: Digital Input/Output connector.

PXT

Connect the digital input signals to the green connectors on the camera module next to the PXT sensor, see Figure 3.

Pin of C1/C2 Name  Description
⏚ Ground  Ground of Start signal
DI1 Start  Measurement Start/Stop (+24 V versus ⏚)